ZOLL AED 3 Battery
The ZOLL AED 3 Battery is designed for use across the ZOLL AED 3 Defibrillator range.
Designed to ensure your ZOLL AED 3 defibrillator can deliver life-saving shock therapy to a casualty, this vital AED battery can deliver up to 140 shocks to a patient without needing to be replaced.
This battery also comes with a shelf life of up to 5 years while performing weekly self-tests on the defibrillator to make sure it remains in rescue-ready condition.
(Please note that AEDs stored outside in low temperatures outside of a thermally controlled enclosure, and any manual tests performed by the device, can impact the lifespan of the battery).
Please Note: The battery model number is 8000-000696. This model is not approved for aviation use.
Key Features
- For use exclusively with the ZOLL AED 3 Defibrillators
- Designed to deliver up to 140 shocks without needing to be replaced
- Up to 5 Year Battery Lifespan (When device performs weekly self-tests)
- Up to 3 Year Battery Lifespan (When device performs daily self-tests)
- Manufacturer: ZOLL
How to Change the Battery on Your ZOLL AED 3
The ZOLL AED 3 runs on a single non-rechargeable lithium battery pack with up to 5 years' shelf life. Swapping it takes a couple of minutes and needs no tools.
Before You Start
- Turn the AED 3 off before you begin
- Check the new battery pack hasn't passed its "install by" date, printed on the label.
- Note the date you're installing the battery and add 5 years to it; that gives you an approximate expiry date to track, separate from the "install by" date on the pack.
Step 1: Turn the unit around and access the battery
With the AED 3 facing away from you, slide the pad pack, the rescue kit sits alongside it, out of its compartment on the back of the unit. This reveals the battery compartment underneath. The pads stay connected by their cable the whole time; you're only sliding the folded packet out of its holder, not unplugging anything.
Step 2: Remove the old battery
Place your fingers into the wells either side of the battery compartment, squeeze the side tabs, and pull the battery straight out. Dispose of it responsibly; the lithium content is harmful if it ends up in general waste. Recycle Now can help you find your nearest battery recycling point.
Step 3: Insert the new battery
Hold the new battery by its side tabs and slide it into the compartment, lining up the notch on the battery with the matching shape in the battery well. Push until it clicks into place.
Step 4: Slide the pad pack back into place
Return the pad pack and rescue kit to its compartment on the back of the unit, the same way it came out.
Step 5: Let the self-test run
The self-test starts automatically the moment the new battery is inserted; you don't need to press anything to prompt it. Once it starts, you'll see a battery icon on the screen, it can occasionally show slightly less than a full charge, since manufacturers test the battery before it ships to you.
Step 6: Confirm the unit is ready
Listen for the “Unit OK” voice prompt and check for the green tick in the status indicator window beside the On/Off button; that confirms the new battery and pads are properly installed, and the AED 3 is ready for service.
Note:
- Standby life is up to 5 years once installed, though frequent self-testing or a poor Wi-Fi signal (if selftest reporting is switched on) can shorten that.
- Never lose track of expiry dates. Register for our free Pads & Battery Reminder service, and we'll email you two months before your ZOLL AED 3 battery is due to expire.
Frequently Asked Questions
Do I need to turn off my ZOLL AED 3 to change the battery?
Yes. If you swap it while the pads are still connected and the unit is in Rescue Mode, it'll power on briefly, show the battery icon, then shut down on its own, just press the On/Off button to restart it normally.
How do I know when to replace my ZOLL AED 3 battery?
Look out for any of these signs:
- The status indicator on the front of the device shows a red X or a blank window instead of the green tick.
- The unit chirps or beeps once a minute while in standby.
- The AED says "Change battery" when you turn it on.
- The install-by or expiration date printed on the battery pack has passed (batteries typically last up to 5 years once installed, depending on how often self-tests run).
We recommend that you replace the battery as soon as you notice any of these, and keep a spare on hand at all times.
How do I dispose of my old ZOLL AED 3 battery?
Don't put it in general waste, it contains lithium. Recycle it at a local collection point; Recycle Now can help you find your nearest one.
How do I work out my ZOLL AED 3 battery's expiry date?
Note the date you install it and add five years; that gives you an approximate expiry date to track alongside the manufacturer's printed “install by” date.
| Brand | Zoll |
|---|
-
Defib Store 4000 Outdoor Defibrillator Cabinet - Keypad Lock - Heater and LED Light - YellowInc.VAT £634.80 £529.00 -
Defib Store 4000 Outdoor Defibrillator Cabinet - Unlocked - Heater and LED Light - GreenInc.VAT £570.00 £475.00 -
Defib Store Bleed Control Cabinet - Unlocked - RedInc.VAT £348.00 £290.00